00:00:00,004 --> 00:00:04,894
Click Open project and open the saved Figure project.

2
00:00:04,894 --> 00:00:14,121
We will measure the 30 kDa GSDMD signal on the lower half of the
membrane, using HSP90 on the upper half as the loading control.

3
00:00:14,121 --> 00:00:19,385
Return to the GSDMD chemiluminescence image.

4
00:00:19,385 --> 00:00:26,495
Drag a horizontal selection around the
lower, 30 kDa signal, covering all six lanes.

5
00:00:26,495 --> 00:00:33,065
Adjust the top and bottom edges to include this
row of bands and a little surrounding background.

6
00:00:33,065 --> 00:00:38,610
Name this cropped band strip GSDMD 30 kDa.

7
00:00:38,610 --> 00:00:41,817
Click Save strip.

8
00:00:41,817 --> 00:00:50,713
The new strip contains only the 30 kDa signal. The
earlier crop with both signals is still in the strip library.

9
00:00:50,713 --> 00:00:56,778
Open Figure builder, then click New figure to create a Figure for quantification.

10
00:00:56,778 --> 00:01:03,302
Name this Figure GSDMD 30 kDa and HSP90 quantification.

11
00:01:03,302 --> 00:01:06,630
Set Lanes to six.

12
00:01:06,630 --> 00:01:12,742
Drag GSDMD 30 kDa from the strip library into the Figure.

13
00:01:12,742 --> 00:01:17,453
Then drag in HSP90 from the same membrane.

14
00:01:17,453 --> 00:01:24,188
Enter the six sample names in matching order: three
Control samples, followed by three Drug samples.

15
00:01:24,188 --> 00:01:29,671
The target and loading control in each column represent the same sample.

16
00:01:29,671 --> 00:01:33,259
Open Protein presets.

17
00:01:33,259 --> 00:01:37,533
Enter HSP90 as the protein name.

18
00:01:37,533 --> 00:01:41,582
Enter 90 as the molecular weight.

19
00:01:41,582 --> 00:01:45,072
Select Loading control.

20
00:01:45,072 --> 00:01:48,189
Click Save entry.

21
00:01:48,189 --> 00:01:51,607
Click Densitometry.

22
00:01:51,607 --> 00:01:56,673
Choose Target / same-membrane control for normalization.

23
00:01:56,673 --> 00:02:03,056
Select HSP90 as the loading control for GSDMD 30 kDa.

24
00:02:03,056 --> 00:02:11,236
These strips come from the upper and lower halves of the
same membrane. Their six lanes are in the same sample order.

25
00:02:11,236 --> 00:02:16,173
For band polarity, choose Dark bands on light background.

26
00:02:16,173 --> 00:02:25,532
Each lane box defines the measurement region. The software estimates
background within the box and sums the background-corrected signal.

27
00:02:25,532 --> 00:02:32,221
First, turn off Equal lane widths and Equal box
heights so you can adjust the boxes individually.

28
00:02:32,221 --> 00:02:37,081
Drag the middle boundary to keep neighboring lanes separate.

29
00:02:37,081 --> 00:02:42,278
Select Equal lane widths to make all six boxes the same width.

30
00:02:42,278 --> 00:02:48,021
Move one lane box vertically to align it with the 30 kDa signal.

31
00:02:48,021 --> 00:02:52,525
Drag the blue edge to adjust the height of this box.

32
00:02:52,525 --> 00:02:57,791
Select Equal box heights to give all six boxes the same height.

33
00:02:57,791 --> 00:03:06,992
Check each lane. Include its signal and a little background,
keeping the upper, 53 kDa signal outside the boxes.

34
00:03:06,992 --> 00:03:11,759
Now look at the six HSP90 lane boxes.

35
00:03:11,759 --> 00:03:16,751
Check that Equal lane widths is selected for HSP90.

36
00:03:16,751 --> 00:03:21,940
Keep Equal box heights selected for HSP90 as well.

37
00:03:21,940 --> 00:03:30,292
The loading control is at a different vertical position.
Check both sets of boxes against the same sample order.

38
00:03:30,292 --> 00:03:35,179
For GSDMD, click Confirm lane regions.

39
00:03:35,179 --> 00:03:40,388
For HSP90, click Confirm lane regions as well.

40
00:03:40,388 --> 00:03:47,962
Click Calculate ratios to measure the original image pixels
and calculate background-corrected integrated density.

41
00:03:47,962 --> 00:03:54,997
The result table now lists the target signal,
loading-control signal, and ratio for all six lanes.

42
00:03:54,997 --> 00:04:04,180
For each lane, the ratio is the integrated density of GSDMD
30 kDa divided by that of the corresponding HSP90 lane.

43
00:04:04,180 --> 00:04:14,429
Check the six sample names, loading-control values, and background before
exporting. These ratios have not been rescaled to a control-group mean of one.

44
00:04:14,429 --> 00:04:20,186
Click Export Excel to save the six lane results and their selection records.

45
00:04:20,186 --> 00:04:25,517
Open the exported Excel workbook to view the results in English.

46
00:04:25,517 --> 00:04:33,231
Each lane includes the target and loading-control integrated
densities, their ratio, and the saved ROI coordinates.

47
00:04:33,231 --> 00:04:36,743
Return to the software.

48
00:04:36,743 --> 00:04:43,187
Click Back up project to save the crop, lane
regions, and loading-control settings together.

49
00:04:43,187 --> 00:04:51,291
When you reopen this project, you can check the original
images and lane boxes, then recalculate the results you need.